SECTION 6: PREVENTATIVE MAINTENANCE
Part No. 715-0080, Rev. B
- 59 -
CLEANING OF INCUBATOR AND CHAMBER
WARNING:
The incubator should be turned off and AC or DC power disconnected
when cleaning.
CAUTION:
CLEANING AND CARE: Do not autoclave, pressure sterilize, or gas sterilize
the incubator, cables, or sensors. Use cleaning solutions sparingly as
excessive solution can flow into the incubator and cause damage to internal
components. Do not soak or immerse the incubator or sensors in any liquid.
Do not use petroleum-based, alcohol, acetone, or other harsh solvents. See
cleaning instructions of reusable sensors and cables in directions for their
use.
Wash the chamber, infant mattress, and top surface of the airflow tray with mild soap or
detergent and water solution.
CAUTION:
Do not drip cleaning solution through the holes where the swell latches fit into
the air flow assembly.
Household ammonia diluted in water in the concentrations recommended for hospital cleaning
may also be used. These solutions may be applied with a soft absorbent cloth, followed by
rinsing with clean water. Wipe away residue with soft chamois or cellulose sponge material.
CAUTION:
The infant tray grounding tabs are sharp; use care when cleaning the air flow
system.
CLEANING OF CABLES (OPTIONAL EQUIPMENT)
Clean the oxygen sensor cable and the pulse oximeter patient cable by wiping with a 70%
isopropyl alcohol pad (or equivalent) and allow the cables to dry before reusing.
CLEANING OF PULSE OXIMETRY SENSOR (OPTIONAL EQUIPMENT)
The provided sensors are disposable and should only be reused on the same patient if the
adhesive still adheres to the skin and the detector window is clear. Otherwise, properly dispose
of the sensor (see SECTION 6, PRODUCT DISPOSAL / RECYCLING, for disposal methods).
For reusable sensors, see the manufacturer’s cleaning instructions.
CLEANING OF OXYGEN SENSOR (OPTIONAL EQUIPMENT)
When cleaning or disinfecting the oxygen sensor, take appropriate care to prevent any solution
from entering the sensor end or the connector end. The oxygen sensor may be cleaned using a
mild detergent and moist cloth and disinfected using standard topical disinfectants.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...