SECTION 5: OXYGEN MONITOR (OPTIONAL FEATURE)
Part No. 715-0080, Rev. B
- 52 -
OXYGEN MONITOR SET-UP INSTRUCTIONS
The following steps will instruct the user how to initially set up the oxygen monitor, program low and
high alarm settings, and how to calibrate the oxygen sensor.
CAUTION:
Check alarm limits each time the system is used to ensure that they are appropriate
for the patient being monitored.
INITIAL SET-UP
1.
Turn the incubator power on using the Power Button (Figure 3 - 1, item 32).
2.
Connect the oxygen sensor cable to the connector on the side of the incubator
(Figure 3 - 2, item 1).
3.
Connect the oxygen sensor to the cable and place in area to be monitored.
a.
Place the oxygen sensor inside the infant chamber to monitor ambient oxygen
concentration within the chamber.
b.
Connect the oxygen sensor to a 15 mm adapter on a patient breathing circuit to
use in conjunction with a ventilator.
4.
To turn on the Oxygen display, press the SET button (Figure 3 - 1, item 24) two times
and then either the up or down arrow buttons (Figure 3 - 1, item 25) until the OXYGEN
screen displays “On”. If the screen shows “---”, there is no value available.
5.
When the Oxygen Monitor is disabled, the OXYGEN screen will display “OFF”. All
alarms related to the Oxygen Monitor will also be disabled.
SETTING HIGH AND LOW OXYGEN ALARMS
NOTE:
The Oxygen Monitor display will convert back to Normal Operation Mode after 5
seconds if no buttons are pressed.
1.
Press the SET button (Figure 3 - 1, item 24) seven times (or until the OXYGEN HI LED
illuminates) to set the high oxygen alarm.
2.
Use the up or down arrow buttons (Figure 3 - 1, item 25) to set the high alarm to the
desired setting.
a.
Must be at least 1% higher than the low oxygen alarm setting.
b.
Can be set from 31 - 100%.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...