SECTION 3: OPERATING INSTRUCTIONS
Part No. 715-0080, Rev. B
- 41 -
When the battery voltage drops below 11 volts, the
LOW BAT
indicator (Figure 3 - 1, item 27)
illuminates and an intermittent audible alarm turns on. The battery will be able to supply the heater
requirements for only a few minutes after this alarm occurs. The
LOW BAT
alarm cannot be reset.
The incubator must be connected to AC power to allow the battery to recharge when this alarm occurs.
NOTE:
When this alarm activates, the incubator has approximately fifteen (15) minutes of
operational power remaining. This estimate is for a good battery and an ambient
temperature of 20
C.
EXTERNAL LIGHTING
The incubator is equipped with an external examination light. This light can be activated only when the
incubator is turned on.
The light is intended to be used as an external light source. Do not place the light into the inner infant
chamber.
WARNING:
Do not change the lamp and touch the patient simultaneously.
WARNING:
EXPLOSION HAZARD: Do not use the examination light, the pulse oximeter,
or oxygen monitor in the presence of flammable anesthetics or other
flammable gases.
WARNING:
Avoid eye exposure. Direct light exposure may cause eye damage. Infants
must wear eye protection.
SKIN TEMPERATURE PROBE
The Measurement Specialties 409B Reusable Skin Temperature Probe is to be used to monitor the
surface temperature of the infant. Select appropriate site for monitoring according to currently
accepted medical practices.
WARNING:
The skin temperature sensor is not a rectal probe. The skin temperature
sensor is not to be used as a rectal probe.
Содержание AirBORNE A750i
Страница 1: ...A750i Transport Incubator Service Manual...
Страница 6: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 4 SCLAIM MER...
Страница 7: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 5 SCLAIM MER...
Страница 8: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 6 SCLAIM MER...
Страница 9: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 7 SCLAIM MER...
Страница 10: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 8 SCLAIM MER...
Страница 11: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 9 SCLAIM MER...
Страница 12: ...Par rt No 715 008 80 Rev B LA ANGUAG GE DIS 10 SCLAIM MER...
Страница 102: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B CONTROL SCHEM L BOARD S MATICS SCHEMAT DRAW TIC 001 51 WINGS 81 100...
Страница 103: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM PULSE MATICS OX DISPLA DRAW AY BOARD WINGS D SCHEMA ATIC 001 5 5193 101...
Страница 104: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 102...
Страница 105: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 103...
Страница 106: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 104...
Страница 107: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 105...
Страница 108: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 106...
Страница 110: ...S Par ECTION rt No 715 008 N 14 S 80 Rev B SCHEM MATICS DRAW WINGS 108...
Страница 112: ...S Par ECTION rt No 715 008 SCHE N 14 S 80 Rev B MATIC PN SCHEM NEUMATIC MATICS C A750i CA DRAW ART 002 40 WINGS 010 110...