EN - ENGLISH
11
INSTRUCTIONS FOR USE
Power On
Power on the Generator by pushing the power button. The power button will illuminate blue. An error in the start-up
indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Self-Test
The Generator performs a self-test after it is powered on. All display elements on the front panel are turned on during this process.
The following light up briefly in the order they are listed: The MONOCUT and BICUT mode buttons, then MONOCOAG and
BICOAG mode buttons, followed by the red indicator light above the neutral electrode receptacle. The two Output Power displays
light up with “88” and the white CUT and COAG Indicator lights illuminate sequentially; the respective activation tone must sound
simultaneously. The default start-up of the device is BICUT and BICOAG with both displays lit up with “00”.
An error in the self-test sequence indicates a device defect. Refer to the Error Diagnosis/ Troubleshooting section of this IFU.
Содержание Codman 901001ESUO
Страница 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 137: ...EL 137 Codman Codman Codman Codman Codman Integra...
Страница 138: ...EL 138 RF HF RF N2O...
Страница 139: ...EL 139 4 5 3 2 3 0 10 C 50 F 10 C 50 F IEC 60601 1 16 70...
Страница 142: ...EL 142 13 14 15 16 17 18 19 13 14 15 16 17 18 19 100 240 V 50 60 Hz T3 15 AH 250V...
Страница 143: ...EL 143 20 21 22 23 20 21 22 CUT 23 COAG...
Страница 144: ...EL 144 1 4 1 2 COAG CUT Codman Electrosurgical Generator W W BICOAG BICUT MONOCUT Monopolar Bipolar MONOCOAG 1 2 Codman...
Страница 145: ...EL 145 MONOCUT BICUT MONOCOAG BICOAG 88 CUT COAG BICUT BICOAG 00...
Страница 158: ...EL 158 Monocut Monocoag Bicut Bicoag Crescendo CF IP2X 12 5 mm 40o C 40o F 60o C 140o F 15 85 1050hPa 500hPa...