Connecting the Neutral Electrode
Connect the Neutral Electrode to the receptacle on the front of the device.
NOTE:
• If the neutral electrode is
NOT
connected during monopolar mode, the “Neutral electrode not
connected” indicator above the receptacle will light up red and flash.
• If the finger switch or Foot Pedal is activated while in monopolar mode without the neutral electrode
connected, an audible warning sounds. RF current cannot be activated. This has no influence on the
bipolar current.
• When using a multi-surface electrode, the “Neutral electrode not connected” indicator will turn off after
the safe application state is reached.
• The entire area of the neutral electrode should be reliably attached to a suitably prepared and appropriate
area of the patient’s body as defined by the manufacturer.
Connecting Monopolar Accessories
Connect a 3-pronged connection monopolar cord or bovie connection monopolar cord to the appropriate location in the monopolar
receptacle. Ensure the 3-pronged connection is in the correct orientation (see illustration below).
Connect the desired Monopolar Instruments to this cord.
NOTE:
Connect a Monopolar device with a finger switch for CUT and COAG activation.
OR
Connect a Monopolar device without a finger switch, in combination with the Foot Pedal for CUT and
COAG activation.
Connecting Bipolar Accessories
Connect a Codman Bipolar cord to the receptacle. See Recommended Accessories section.
Connect the desired bipolar instrument to the cord. The Foot Pedal is used for CUT and COAG activation.
NOTE:
• The neutral electrode is
NOT
required for bipolar applications.
Monopolar
Bipolar
INSTRUCTIONS FOR USE: CONNECTING ACCESSORIES
1
1
1
2
2
Bipolar
a1
a2
Содержание Codman 901001ESUO
Страница 2: ...2 THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 137: ...EL 137 Codman Codman Codman Codman Codman Integra...
Страница 138: ...EL 138 RF HF RF N2O...
Страница 139: ...EL 139 4 5 3 2 3 0 10 C 50 F 10 C 50 F IEC 60601 1 16 70...
Страница 142: ...EL 142 13 14 15 16 17 18 19 13 14 15 16 17 18 19 100 240 V 50 60 Hz T3 15 AH 250V...
Страница 143: ...EL 143 20 21 22 23 20 21 22 CUT 23 COAG...
Страница 144: ...EL 144 1 4 1 2 COAG CUT Codman Electrosurgical Generator W W BICOAG BICUT MONOCUT Monopolar Bipolar MONOCOAG 1 2 Codman...
Страница 145: ...EL 145 MONOCUT BICUT MONOCOAG BICOAG 88 CUT COAG BICUT BICOAG 00...
Страница 158: ...EL 158 Monocut Monocoag Bicut Bicoag Crescendo CF IP2X 12 5 mm 40o C 40o F 60o C 140o F 15 85 1050hPa 500hPa...