3) Set the mode of the voltmeter to AC voltage measurement.
4) Start the operation by K2.
5) Next, set the indicated value of the voltmeter to “the calculated value of 1)
±0.5mV” by adjusting the drive voltage of K2. Record the value of the adjusted
drive voltage to a check sheet.
6) Record these items indicated on the display ;
ACC (acceleration) [m/s
2
]
7) Change the connection of [OUTPUT Ch1] to [INPUT Ch2] of the I/O UNIT. Also
change the input setting of K2 to Ch.2 from Ch.1. The input sensitivity of Ch.2 is
the same as Ch.1.
8) Operate the same procedure as 3)
~
6), and record the measured data.
9) These procedures are needed to be operated also for Ch.3 and Ch.4. When the
optional input channels are added, the input channels Ch1. ~ Ch8 of the expansion
board are also needed to be measured and recorded.
10) Change the input type to [Charge input (1mV/pC)] from [Charge input
(10mV/pC)] . Operate the same procedure as [Charge input (1mV/pC)] and record
the measured data.
Criterion
Accelerometer : 100.00m/s
2
within ±2
%
(98.00 ~ 102.00)
Monitoring output : input voltage within ±5
%
(335.825
~
371.175)
<
Check Sheet
>
For conducted maintenance items next page, put a circle in the “Result” column.
If the system has a fault or damage, or if a defective part is replaced, enter the description
in the “Remarks” column.
After checking all maintenance items, send this check sheet to us by E-mail.
Содержание K2-CE
Страница 1: ...Vibration Control System K2 CE K2Sprint CE Common Part TYPE II Instruction Manual IMV CORPORATION...
Страница 4: ......
Страница 34: ...2 6 Step 2 The initial screen of installation appears Press the button of Next...
Страница 38: ...2 10 Step 2 The initial screen of installation appears Press the button of Next...
Страница 44: ...2 16 Step 8 The screen of finishing the device driver installation appears Press the button of Finish...
Страница 49: ...2 21 Step 2 The initial screen of installation appears Press the button of Next...
Страница 55: ...2 27 Step 3 The screen displaying the message of Uninstall or change a program appears...
Страница 56: ...2 28 2 Windows 10 Procedures Step 1 Right click the Start Menu And select Control Panel on the displayed menu...
Страница 58: ...2 30 Step 3 The screen displaying the message of Uninstall or change a program appears...
Страница 81: ......
Страница 82: ......
Страница 92: ...4 3 4 2 2 Exit from Application 1 K2 Launcher Press the button Close on the screen...
Страница 107: ...4 18 Step 3 A graph of the page displayed currently appears in the display...
Страница 110: ...4 21 Step 3 Additional graph appears on the currently displayed page...
Страница 115: ...4 26 Step 2 Change values of the graph range for horizontal axis displayed and press the OK button...
Страница 121: ...4 32 Step3 Press the Mark button to mark the cursor data displayed in the list...
Страница 141: ...4 52 Step 2 Press Yes when a dialog of message for assuring appears...
Страница 148: ...4 59 Step 3 Press the Define button of Contact I O information Step 4 Operate the setting of Contact I O and press OK...
Страница 149: ...4 60 Step 5 Press the OK button...
Страница 170: ...4 81 Step 9 Press OK after completing of the necessary settings...
Страница 172: ...4 83 Report of testing result generated by Report Generator...
Страница 184: ...5 6 5 4 1 Initial setting factory setting Factory setting is as shown below To restore it click Initialization...
Страница 189: ...6 5 6 2 2 Display of Standard Graph Procedure Step 1 Press the Open button Step 2 Select the graph data file...
Страница 196: ...6 12 Step 3 Select desired data file Some files can be selected at a time...
Страница 198: ...6 14 Step 5 Graph window is added and the selected graph appears Change of displayed graph Press the graph change button...
Страница 201: ...6 17 Step 3 Select desired data file Some files can be selected at a time...
Страница 212: ...7 6 7 2 2 Standards Mode Procedures Step 1 Press the button of Standards...
Страница 217: ...7 11 7 2 3 Test files mode Procedures Step 1 Press the button of Test files...
Страница 257: ...Waterfall graph depth fixing 4 19 Waterfall graph scroll display 4 19 4 22 4 27 Web Monitor 4 66...