17
Troubleshooting E2 Error Code
Error Code E2 indicates a timeout before the chamber is filled with water.
1.
Remove left and right side covers from unit.
2.
Check flow switch on right side of unit. If cleaning solution is leaking around wires of flow
switch disconnect flow switch wires and start a P0 cycle. If unit fills with water and fails
with an ED error code replace flow switch (#01-111568S).
3.
Check for water in the bottom pan by the float. If water is in the bottom pan siphon water
out of pan and start a P0 cycle. Sidebag should fill with cold water through water inlet hose.
If sidebag does not fill check cold-water inlet valve. If sidebag fills water will rise to red
line at top of sidebag then flow over and down channel to emitter housing. When the
emitter housing is full the pressure switch will activate and the blue arm will be pushed out
activating the electrical switch. If the blue arm does not activate the electrical switch
replace the emitter housing (#13447.01 Red or #13447.00 Clear). When the electrical
switch activates the drain solenoid should open and the sidebag should empty into the
chamber. The blue arm should go back in once the drain valve opens. If the electrical
switch activates check for line voltage (208-240VAC) to the drain solenoid (#13614.00). If
the drain solenoid needs to be replaced and the sidebag is full of water loosen the 2 screws
holding the drain solenoid,
do not remove them
, and pull out on the drain solenoid until
the water in the sidebag flows by the solenoid and into the chamber.
Bottom Pan Float
Side bag
Inlet Hose
Red Line
Emitter Housing Red
Blue Arm
Drain Valve