13
Hydrim L110W P3 & P4 Cycle with Revision 2.02 Software (Cond.)
9.
Fill wash chamber target temperature 50
°
C/122
°
F
When sidebag is full and pressure switch B1 activates turn ON Y6 sidebag drain valve to
allow sidebag to drain into the chamber. Pressure switch will deactivate when sidebag starts
draining into chamber. After approximately 45 seconds turn ON circulation pump M1 and
Y7 hot water valve. The on time for the hot water inlet valve depends on the “Water Debit
Constant” that was calculated previously. The chamber capacity is 4 liters, 2.7 liters from
the sidebag and 1.3 liters from the hot water inlet valve, resulting valve Y6 is on for
approximately 75 seconds and valve Y7 is on for 1.3/Water Debit Constant. When water
inlet valves timer expires add 10 seconds to the sidebag drain valve Y6 before closing.
Failure to fill chamber will cause an
E2 code
.
10.
Circulation, heating and dosing target temperature 50
°
C/122
°
F
Start circulation pump M1, dosing pump M4 and turn ON water heater R1 if water
temperature in chamber is below target temperature. Start heating and dosing error timers.
Dosing pump runs for 7 seconds. If flow switch does not detect dosing before the dosing
error timer ends (30 seconds), the Hydrim fails with an
ED code
. Once selected dosing time
is achieved the dosing pump will turn OFF.
11.
Circulation and heating target temperature 50
°
C/122
°
F
Circulation pump continues to run and water heater remains turned ON. If the heating error
timer ends and the chamber water temperature is below the target temperature, the Hydrim
fails with an
E1 code.
12.
Circulation, heating and dosing
Circulation pump continues to run, and water heater remains turned ON. When water
temperature in chamber reaches target temperature start dosing error timer. The dosing
pump runs for 7 seconds. If flow switch does not detect dosing before the dosing error timer
ends (30 seconds), the Hydrim fails with an
ED code
. Once selected dosing time is
achieved the dosing pump will turn OFF.
13. Circulation
Circulation pump M1 remains ON for 360 seconds/6 minutes.
14.
Circulation and evacuation 10 seconds
Circulation pump M1 remains ON. Turn ON drain pump M2.
15.
Evacuation 20 seconds
Circulation pump M1 turns OFF. Drain pump M2 remains ON for 20 seconds.
16.
Fill wash chamber target temperature 40
°
C/104
°
F without filling sidebag
Turn ON valve Y6 and valve Y1 or Y7 (depending on water inlet temperature and the target
temperature). The ON time for the water inlet valve depends on the “Water Debit Constant”
that was calculated previously. Turn valve Y1 or Y7 ON for 3/Water Debit Constant and
Y6 for 3/Water Debit Co 10 seconds. When water inlet valves timer expires add 10
seconds to the sidebag drain valve Y6.
17.
Evacuation 30 seconds fill sidebag
Drain pump remains ON and sidebag starts filling target temperature 45
°
C/113
°
F.
18.
Wait until sidebag is full target temperature 45
°
C/113
°
F
19.
Fill wash chamber target temperature 45
°
C/113
°
F
20.
Circulation 60 seconds/1 minute target temperature 45
°
C/113
°
F fill sidebag
21.
Circulation and evacuation 10 seconds
22.
Evacuation 20 seconds
23.
Wait until sidebag is full target temperature 80
°
C/176
°
F
24.
Fill wash chamber target temperature 80
°
C/176
°
F