11
Hydrim L110W P1 & P2 Cycle with Revision 2.02 Software (Cond.)
11.
Evacuation 30 seconds fill sidebag target temperature 45
°
C/113
°
F
Drain pump remains ON and sidebag starts filling target temperature 45
°
C/113
°
F.
12.
Wait until sidebag is full target temperature 45
°
C/113
°
F
13.
Fill wash chamber target temperature 45
°
C/113
°
F
14.
Circulation 60 seconds/1 minute target temperature 45
°
C/113
°
F fill sidebag
15.
Circulation and evacuation 10 seconds
16.
Evacuation 20 seconds
17.
Wait until sidebag is full target temperature 80
°
C/176
°
F
18.
Fill wash chamber target temperature 80
°
C/176
°
F
19.
Circulation, heating and dosing
Start circulation pump M1, dosing pump M4 and turn ON water heater R1 if water
temperature in chamber is below target temperature. Start heating and dosing error timers.
Dosing pump runs for 5 seconds. If flow switch does not detect dosing before the dosing
error timer ends (30 seconds), the Hydrim fails with an
ED code
. Once selected dosing time
is achieved the dosing pump will turn OFF.
20.
Circulation and heating target temperature 65
°
C/149
°
F
Circulation pump continues to run and water heater remains turned ON. If the heating error
timer ends and the chamber water temperature is below the target temperature, the Hydrim
fails with an
E1 code.
21.
Circulation 180 seconds/3 minutes
22.
Circulation and evacuation 10 seconds
23.
Evacuation 20 seconds
24.
Drying (P2 cycle only)
Dryer time factory set to 10 minutes. May be increased to 20 minutes in 1-minute
increments. Dryer will run at slower speed for 1 minute then increase to high speed.