Troubleshooting E4 Error Code
Error Code 4 indicates failure of water evacuation from chamber before 60-second timeout.
1.
If the chamber is empty remove the filters from inside the chamber and pour some water
into the base of the chamber. Turn the unit ON and select P0. The drain pump should turn
ON and evacuate the water from the chamber. If you do not hear the drain pump running
call SciCan for technical support. If the drain pump runs but the water does not evacuate
proceed to step 2.
2.
Remove the left side panel from the unit.
3.
Check for debris in the sidebag (#13462.00) drain channels between the two drain tube
connections. Remove both metal mounting brackets to check channel behind brackets. If
channel is clean proceed to step 4.
4.
Remove hose from drain pump where it connects to the side bag. Place hose in a bucket.
Remove the filters from inside the chamber and pour some water into the base of the
chamber. Turn the unit ON and select P0. The water in the chamber should pump into the
bucket. If water does not evacuate from chamber call SciCan for technical support.
Hose from Drain Pump
Side Bag
Drain Hose to back of Unit
Mounting Brackets
Drain Channel In
Drain Channel Out