Connect valve plate by connecting 2 center screws first, then 3 corner screws using 8mm nut driver.
Screws are self-tapping and corner screws need to thread into the plastic.
Removal and Installation of Flow Switch Assembly with Tubing
Remove the old flow switch and tubing going to
Attach new flow switch wires to terminal block
dosing pump and chamber. Retain tubing attached and connect ground wire Neatly tie wrap wiring
bottom of flow switch and attach to new switch.
to bracket. Connect new gray tubing to dosing pump and
secure with a cable tie.
Screw the flow switch bracket to the unit bracket Bolt the bracket at the base of the unit by
using M4 screw provided.
removing the nut and washer.
Flow Switch
Tubing
Tubing
New gray tubing
Nut & washer
Center Screws