HYDAC FILTER SYSTEMS GMBH
Operation | 6
BeWa CTU1x5x 4603539 en-us web
57 / 120
Fig. 23:
Extraction chamber with basket mount
The dimensions of the hook-in basket are 220 x 110 x 40 mm.
Mesh width = 4 mm.
Proceed as follows for spray extraction:
ü
Determine the negative control value; for details, see
mining the negative control value - Settings/parameters
Flush the inner chamber including the auxiliary equipment
such as the hook-in basket, grating, mounts for the test item,
etc.
1. Place the test item to be sampled in the extraction chamber.
2. Close the door and lock it.
3. Select the required test parameters on the user interface and
start the spray extraction.
4. Spray the component surfaces to be analyzed evenly.
The test liquid, loaded with the partially rinsed off particulate
contamination, is drawn across the filter membrane using a
vacuum.
ð
Wait until the test liquid has been fully suctioned through the
filter membrane.
1
2
3
4
5
6
7
8
9
A