5.2.2 Ultrasound unit / mounting/connecting the vibrator
(optional)
This chapter contains information on the installation and connec-
tion of the ultrasound generator and the ultrasonic vibrator.
1. Place the ultrasound generator in the switch cabinet at the
bottom let in the corner.
2. Connect the electric cable according to the following repre-
sentation.
1
2
5
4
3
a
b
Fig. 16:
Connecting the ultrasonic unit/vibrator
1 Ultrasonic converter
2 Ultrasound generator
3 Ultrasound generator plug <-> Ultrasonic converter
4 Mating connector XD3 in the switch cabinet
5 Extraction chamber
a High frequency cable
b Supply voltage for the ultrasound generator
O
The components are connected and operational.
5.3 Suctioning the machine room
The back of the extraction unit has a suction connection of
Ø 100 mm for the machine room.
For this, follow the safety data sheet of the test liquid being used.
It is advisable to suction the machine room if:
– the chemical composition of the test liquid used specifies it
– there is no ventilation and/or air conditioning in the operating
room of the CTU
– an easily volatile test liquid is used
Details on this can be found in the
of the test liq-
uid.