---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 85
(6) Eyeball Diagram representing refraction power
It is possible to print the refraction state diagram regarding the result of the REF measur
ement to the internal printer. This function will be performed when ‘
PRINT=EYE: ON
’ in t
he Page 3/4 of the SETUP MODE. See the following table and diagram for further infor
mation.
Classification of Data
NORMAL
MYOPIA HYPE
ROPIA
ASTIGM
ATISM
NORMAL
-0.5 =<S<= +0.
5
-0.5 =<C<= 0.
0
◎
MYOPIA
S< -0.5
-0.5 =<C<= 0.
0
◎
SIMPLE MYOPI
C ASTIGMATIS
M
-0.5 =<S<= 0.0
C < -0.5
◎
◎
COMPOUND M
YOPIC ASTIGM
ATISM
S < -0.5
C < -0.5
◎
◎
MIXED ASTIGM
ATISM
0.0 < S
C < -0.5
S+C < 0.0
◎
HYPEROPIA
+0.5 < S
-0.5 =<C<= 0.
0
0.0 < S+C
◎
SIMPLE HYPE
ROPIC ASTIGM
ATISM
+0.5 < S
C < - 0.5
0.0 =<S+C<= +0.
5
◎
◎
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...