46 Auto Ref/Keratometer CRK-1 ------------------------------------------------------------------------------------------------
[On-focus calibration]
The first thing you have to do is on-focus calibration.
On-focus calibration is for yellow points on KER data graph.
1) Set KER 0D model eye on the chin rest and enter the KER TEST MODE.
2) Enter the KER signal clip test mode.