---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 25
3.3.1.4. Description of wire harness connection of MAINBOARD
Connect No.
Description
Target connector
1 (CCN1)
REF CMOS CAMERA
AR1030 CAMERA BOARD
2 (CN17)
KER CMOS CAMERA
COLORCMOS CAMERA BOAR
D
3 (CN18)
LCD INTERFACE
LCD ASSY
4 (CN19)
RGB OUT
RGB OUT HARNESS
5 (CN16)
READY LED
READY LED HARNESS
6 (CN9)
OPTIC MOVEMENT MOTOR
OM MOTOR HARNESS
7 (CN10)
IR FILTER MOTOR
IR FILTER MOTOR HARNESS
8 (CN14)
UART0 PORT
UART HARNESS
9 (CN6)
UART2 PORT
UART HARNESS
10 (CN24)
JOY STICK
JOYSTICK ASSY
-
Joy-Link HANRESS
-
CRK ENCODER BOARD
11 (CN8)
THERMAL PRINTER
THERMAL PRINTER ASSY
12 (CN20)
Y MOTOR PI
Y MOTOR PI BOARD
13 (CN23)
OPTIC PI ( OM , IR FILTER )
OM , IR FILTER PI BOARD
14 (CN21)
RL ENCODER, PI
RL ENCODER BAORD
RL PI BOARD
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...