76 Auto Ref/Keratometer CRK-1 ------------------------------------------------------------------------------------------------
6. Troubleshooting
Phenomenon 1: When Power is ON, there is no action
Phenomenon 2: LCD monitor display nothing (But other functions work)
Check whether power cable is inserted well
Volt age O
K?
Check fuse in the bottom side
Take off the head cover
1. Check whether the cable from SMPS to
Main Board is inserted well
2. Check the v12V) of connector b
y using DMM, CN1 Pin No.6
1. Take off the bottom case
2. Check whether the power voltage is
No
Yes
IS Voltage Correc
t?
Replace Mainboard
In this case, you should do model
eye setup process again
No
Yes
Replace SMPS
Check Replace Fuse
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...