58 Auto Ref/Keratometer CRK-1 ------------------------------------------------------------------------------------------------
e screen.
2) The machine optical head moved to about the middle position, record scale valu
e.
(
e.g.The screen is displayed at 8cm.
),
click ”CENT” and “SETUP”. The focu
s is adjusted on the screen and prees joystick button.
3) Move the machine optical head 4 cm to the left.The focus is adjusted on the sc
reen and prees joystick button.
4) Move the machine optical head 1 cm to the right.The focus is adjusted on the s
creen and prees joystick button.
(
A total of nine measurements
)
5) After calibration is completed, click
“
SAVE
”
.
D. After PD Setup, engineer mode by pressing the SAVE SYS DATA is stored.
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...