---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 43
4.1.5. Check KER calibration data
If KER signal has no apparent problem, you can check calibration data.
Checking KER calibration data is available at “1-4. KER DATA SETUP”.
Like REF calibration data, ascending or descending current is important.
If there are any points stand against the current, you can say the calibration data is inva
lid.