---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 65
4.8. IQC mode
1) Test procedure
-
Start IQC Mode in the engineer mode.
-
Test the each mode.
-
If Test is success or fails, select pass or fail on pass and fail button.
2) Menu
-
NO DISP
:
All of the sections disappear on the screen.
-
DISP
:
All of the sections appear on the screen.
-
RESET
:
Pass and fail data of all of the sections are reset.
-
REPORT