---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 33
The principle of Retro-Illumination observation
3.5.1.1. Principle of retro-illumination observation
The incident lights of the infrared LED (spot light) onto the fundus of the eye are
retro-scattered from it. The scattered lights from the fundus illuminate the vitreou
s humor and some of it go through the crystalline lens and the iris. If it is captur
ed by TV camera, any turbid part of the transparent vitreous humor, lens and cor
nea can be observed.
3.5.1.2. Usage of Retro-Illumination observation
The turbid part of the cataract and the frontal glass can be
observed and the change of it can be taken up with. Therefore, the reason of the
degradation of eyesight can be known whether it is caused by the simple refractiv
e abnormality or the advance of the cataract.
n
=
1.3375, 1.336, 1.332
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...