80 Auto Ref/Keratometer CRK-1 ------------------------------------------------------------------------------------------------
Phenomenon 6: Cannot see the eye fixation target
Take off the head cover
Take off the optic cover
Check whether led is out
Is LED out?
Replace Mainboard
Replace LED board
No
Yes