---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 59
4.4. Factory Data Setup
1) MAKE SYSTEM BACKUP
-
It is for our factory worker.
-
Please do not touch it.
2) VERIFY BACKUP DATA
-
Check if the valid system backup is stored.
3) RESTORE SYSTEM
-
Restore all system data to factory backup.
4) COPY FRAM