---------------------------------------------------------------------------------------------- Auto Ref/Keratometer CRK-1 57
A. Engineer mode of the PD Setup begins.
B. PD Setup in the menu is as follows.
1) SAVE
- Changed the temporary storage of PD are set.
2) CENT
- The current value of the X-axis encoder center position.
3) + STEP
- The offset of the slit spacing (Slit pitch Offset) increases.
4) - STEP
- The offset of the slit spacing (Slit pitch Offset) decreases.
5) EXIT
- PD Setup will exit.
Slit pitch Offset is changed, the screen update interval of the slit is measured from
which pupil distance (PD) is also changed.
C. PD measurements in the following order, depending on setting.
1) PD measurements and those mounted on the stand, the focus is adjusted on th
Содержание CRK-1
Страница 1: ...Service Manual Auto Ref Keratometer CRK 1...
Страница 18: ...Auto Ref Keratometer CRK 1 17 Use signal amplifier if the distance with the outside monitor is significant...
Страница 19: ...18 Auto Ref Keratometer CRK 1 3 2 Optical System...
Страница 35: ...34 Auto Ref Keratometer CRK 1...
Страница 41: ...40 Auto Ref Keratometer CRK 1...
Страница 45: ...44 Auto Ref Keratometer CRK 1 In that case you should take a calibration again...
Страница 48: ...Auto Ref Keratometer CRK 1 47 3 Control working distance to match HR 4 Touch SETUP button...
Страница 49: ...48 Auto Ref Keratometer CRK 1 5 Touch 0D target 6 Set real target diopter and touch START SETUP button...
Страница 53: ...52 Auto Ref Keratometer CRK 1 Save After calibration you have to save it to FRAM...
Страница 63: ...62 Auto Ref Keratometer CRK 1 4 Touch APPLY button to apply the result of calibration...
Страница 73: ...72 Auto Ref Keratometer CRK 1 2 Point for Assembly Assembly is the reverse procedure of disassembly...
Страница 76: ...Auto Ref Keratometer CRK 1 75...
Страница 88: ...Auto Ref Keratometer CRK 1 87 MATISM Diagram of the eye ball output by REF Measurement...