54 Auto Ref/Keratometer CRK-1 ------------------------------------------------------------------------------------------------
1) Start with default value.
2) Test tracking performance.
A. Move focus cross to outer mire ring edge.
B. Leave the joystick out, and watch the tracking action.
C. If the cross passed the center of mire ring then you can say that the test is fail
ed.