sat 805 Pulse Oximeter
Disinfecting, Cleaning and Maintenance
93
10.3.3
Cleaning of Patient Cables
The patient cable is to be cleaned as follows:
1.
Disconnect from the patient and also from the patient sensor.
2.
Use a cloth rinsed in 70% isopropyl alcohol to wipe the entire patient cable.
3.
Let the entire patient cable completely air dry long enough before reusing.
10.4
Maintenance
User maintenance is not required.
The device itself requires maintenance once every 36 months in order to work within the guaranteed
specifications.
ATTENTION! Unit can be damaged!
Never correct defects on the unit, repair it or perform maintenance yourself! Corrections of
defects, repairs or any maintenance should be done exclusively by HUM Gesellschaft für Homecare
und Medizintechnik mbH or by a HUM Gesellschaft für Homecare und Medizintechnik mbH
authorized medical device supplier!
NOTICE!
A functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a pulse
oximeter
.
NOTICE!
If there is independent demonstration that a particular calibration curve is accurate for the
combination of a pulse oximeter and a pulse oximeter probe, then a functional tester can measure the
contribution of a monitor to the total error of a monitor/probe system. The functional tester can then
measure how accurately a particular pulse oximeter is reproducing that calibration curve
.