sat 805 Pulse Oximeter
Operation
36
6.6.1
Selecting the (correct) Sensor
Select the sensor based on the patient’s age and weight and its re
-usability. The Table is a guide. The
correct sensor has to be matched individually to the patient.
NOTICE!
The latest list of all sensors is directly available from HUM Gesellschaft für Homecare und
Medizintechnik mbH.
6.6.1.1
Adhesive Sensors
p/n
Sensor Type
Patient Type and Weight
Application Site
HPO04-ADTX
LNCS® Adtx
Adults > 30 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO04-PDTX
LNCS® Pdtx
Children and Adults > 10 kg
and < 50 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO04-INF
LNCS® Inf-L
Small Children > 3 kg < 20 kg
Great Toe or Thumb
HPO04-NEO
LNCS® Neo
Newborn, < 3 kg, adults > 40 kg
Foot, alternatively Back of the
Hand
HPO04-NEOPT
LNCS® NeoPt
Neonatal Preterms < 1 kg
Foot, alternatively Back of the
Hand
HPO05-ADT
RD SET Adt
Adults > 30 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO05-PDT
RD SET Pdt
Children and Adults > 10 kg
and < 50 kg
Ring- or Middle Finger of the
non-dominant Hand
HPO05-INF
RD SET Inf
Small Children > 3 kg < 20 kg
Great Toe or Thumb
HPO05-NEO
RD SET Neo
Newborn, Children, Adults < 3 kg
or > 40 kg
Foot, alternatively Back of the
Hand
HPO05-NEOPT
RD SET NeoPt
Neonatal Preterms < 1 kg
Foot, alternatively Back of the
Hand