sat 805 Pulse Oximeter
Operation
35
WARNING! Connectors can disconnect during Movement!
When moving the unit during monitoring, absolutely ensure the correct position of the patient
connector. Should the connector accidentally become disconnected an alarm sounds.
6.6
Using Masimo-Sensors
CAUTION! Danger from not Observing User Instructions!
These sensors were developed only for usage of pulse oximeters with MasimoSET-technology.
The responsible organisation and/or the operator has to
check the compability of the pulse
oximeter, cabel and sensor before every usage, otherwise this could cause dangerous injuries of
the patientPay attention to the notices in the user instruction manual for Masimo sensors. If not
observed the measurement results can be false.
Only use approved Masimo sensors with the unit. Use of other sensors can reduce the
performance capacity of the unit and thus produce danger to the patient!
Tissue damage can be caused by incorrect application or use of a sensor, for example by wrapping
the sensor too tightly. Inspect the sensor site as directed in the Instructions for Use to ensure skin
integrity and correct positioning and adhesion of the sensor.
Do not use damaged sensors or cables!
Do not use a sensor with exposed optical components.
The sensors as well as the patient cable connectors are not waterproof. Never immerse the
sensors or connecting elements in liquid or hold under running water.
Do not sterilize sensors and patient cable by irradiation, steam or ethylene oxide.
NOTICE!
More information on the use of Masimo sensors can be found in the user instructions of the
applicable sensor.
NOTICE!
A complete list of validated Masimo sensors will be provided on request by the manufacturer.