sat 805 Pulse Oximeter
Alarms
76
8.2.1.3
Sensor Alarms
NOTICE!
X-CAL or lifetime related alarms (*) might not be present in all devices.
To confirm, please check in the product info menu (see chapter 7.7.6): only devices with DSP version
5.X.X.X will support the X-CAL related alarms and messages.
Status Line Message
Description
Solution
No sensor
connected!
The sensor is not or incorrectly connected to the
unit.
Check connection between
patient cable and sensor. If
needed, replace sensor.
No adh. sensor
connected!(*)
The sensor is not or incorrectly connected to the
unit.
Check connection between
patient cable and sensor. If
needed, replace sensor.
No cable
connected! (*)
The patient cable is not or incorrectly connected
to the unit.
Check patient cable
connection between unit
and sensor. If needed,
replace the patient cable.
Sensor off
patient!
The sensor is connected to the unit, the unit is
on, but no patient can be recognized.
Position the sensor
correctly.
Defective
sensor!
The device has determined that the sensor does
not function at all or only to a limited extent.
Replace the sensor with a
new Masimo sensor!
Unrecognized
sensor!
The device has detected a sensor that is not
approved for this system
or the lifetime has been exceeded.
Only connect Masimo
sensors
or connect a new sensor!
"Defective adh.
sensor!" (*)
The device has determined that the sensor does
not function at all or only to a limited extent.
Replace the sensor with a
new Masimo sensor!
"Defective
cable!" (*)
The device has determined that the patient
cable does not function at all or only to a limited
extent.
Replace the sensor with a
new Masimo patient cable!