
sat 805 Pulse Oximeter
Safety
12
2.1.1
Possible Misapplications
WARNING! Fire Hazard!
The pulse oximeter may not be used during nuclear magnetic resonance imaging tests. The
electrical current induced there can cause burns. This may affect the MRI image and the MRI
device may affect the accuracy of the Pulse Oximetry parameters and measurements.
–
This device is to be operated by qualified personnel only. This manual, all precautionary information
and specifications should be read before use.
–
The pulse oximeter may not be used to monitor breathing (apnea monitor.). This may take place
only with special respiratory monitoring equipment.
–
A pulse oximeter should be considered an early warning device. As a trend towards patient
hypoxemia is indicated, blood samples should be analyzed by a laboratory co-oximeter to completely
understand the patient’s condition.
–
Pulse rate measurement is based on the optical detection of a peripheral flow pulse and therefore
may not detect certain arrhythmias. The pulse oximeter should not be used as a replacement or
substitute for ECG based arrhythmia analysis.
–
Explosion hazard. The pulse oximeter may not be used in a combustible atmosphere. This can
develop when working with flammable anesthetics, laughing gas, or other combustible gases and
liquids.
–
If an alarm condition (other than the exceptions listed herein) occurs while the alarm tone volume is
set to off, the only alarm indications will be visual displays and symbols related to the alarm
condition.
–
Do not use any extensions power cords or adapters of any type. The plug-in power supply must be
intact and undamaged.
–
The pulse oximeter may not be operated in a switchable power socket. Such a plug socket is
unsuitable for a secured power supply.
–
The pulse oximeter works with an optical measurement procedure and should not be operated near
to strong and/or direct light sources.
–
Only use one sensor per measuring point to avoid interferences.
–
Pay attention to a secured and fixed location of the sensors!