sat 805 Pulse Oximeter
Control
65
7.7.5
External Alarm (
Nurse Call
)
WARNING! Danger by incorrect Configuration!
This setting is only allowed to be modified by experienced and training personal. A wrong setting
might de-activate the external alarm.
If the unit shall be connected via the external alarm interface, this interface can be set in accordance to the
respective system. For this, the relay contact can be configured as "
Norm.(ally) open",
"
Norm.(ally) closed
" or "
Remote Alarm
".
–
In the
Norm. open
setting the contact is permanently closed as long as the alarm condition
continues.
–
In the
Norm. close
setting the contact is permanently open as long as the alarm condition
continues.
–
Before connecting a remote alarm device, you must select the
Remote alarm
setting. This gives a
clocked transmission of coded alarm information from the device to the remote alarm device.
1.
Starting from the main menu, call:
Configuration
Nurse
call
2.
Select the desired setting using the trim
knob (
Norm.open
/
Norm.close
/
Remote Alarm
).
3.
Confirm this setting with
Return
.
Fig. 66: Nurse call Interface Setting
7.7.6
Device Information (
Product Info)
By selecting menu
ProductInfo
you can call up the most important device settings on to the display:
–
Device S/N and software version
–
Version of the Masimo circuit board and its product identification
–
DSP and microcontroller-versions