sat 805 Pulse Oximeter
Alarms
78
Action for System Alarm Occurrence:
1.
Secure monitoring of patient by other means.
2.
Turn off unit by pressing the ON/OFF button.
3.
Turn the unit back ON.
If the self-test occurs without errors, the unit can be used again for monitoring.
If an error occurs during the self-test, turn the unit OFF and take to a medical supplier for repairs.
WARNING! Danger to the Patient!
Never use a unit that indicates a system alarm.
System Alarm Messages:
Message
Source
SYST ALARM: bat low!
The battery voltage is below a level that guarantees
reliable operation.
SYST ALARM: MS-FIFO!
An error has arisen in the Masimo input data memory.
SYST ALARM: MS-timeout!
The Masimo-circuit board has not responded to
several queries or does not provide data for more
than 60 seconds.
SYST ALARM: MS-comm.!
The Masimo-data transfer has a permanent defect.
SYST ALARM: MS-error!
The Masimo circuit board has reported an irreparable
defect.
SYST ALARM: Unexp.reset!
An unexpected device restart has occurred.
SYST ALARM: NVRAM error!
The internal non-volatile memory has reported an
error.
SYST ALARM: WDT error!
The integrated circuit that monitors the sat 805 has
reported an error.
SYST ALARM: Reset monitor
The integrated circuit that monitors the sat 805 has