sat 805 Pulse Oximeter
Operation
31
6.3.1
Mains Power Failure Alarm
In case of a mains power failure, an alarm is triggered respectively. At the same time, the unit automatically
switches from mains to battery supply without any interruption.
1.
Approx. 5 sec. After mains failure, an
alarm (medium priority is triggered).
Fig. 11: Mains Power Failure Alarm
2.
The mains power failure will generate a
corresponding entry into the alarm list.
3.
After 30 seconds, the alarm will be
cancelled. The unit is now in battery
operation mode.
4.
If the mains power supply is re-installed
in the course of the a.m. 30 seconds,
the mains power alarm will be
automatically reset.
NOTICE!
The Mains Power Failure condition is
displayed by the dark mains power LEDs.
Fig. 12: Mains Power Failure in Alarm List