prog.book : chapter1.fm 20 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-20
HP E4418B/E4419B Programming Guide
Using the CONFigure Command
When you execute this command, the power meter presets the best
settings for the requested configuration (like the
MEASure?
query).
However, the measurement is not automatically started and you can
change measurement parameters before making measurements. This
allows you to incrementally change the power meter’s configuration from
the preset conditions. The power meter offers a variety of low-level
commands in the
SENSe
,
CALCulate
, and
TRIGger
subsystems. For
example, if you want to change the averaging use the
[SENSe[1]]|SENSe2:AVERage:COUNt
command.
Use the
INITiate
or
READ?
query to initiate the measurement.
Using READ?
CONFigure
does not take the measurement. One method of obtaining a
result is to use the
READ?
query. The
READ?
query takes the measurement
using the parameters set by the
CONFigure
command then sends the
reading to the output buffer. Using the
READ?
query will obtain new data.
Using INITiate and FETCh?
CONFigure
does not take the measurement. One method of obtaining the
result is to use the
INITiate
and
FETCh?
commands. The
INITiate
command causes the measurement to be taken. The
FETCh?
query
retrieves a reading when the measurement is complete, and sends the
reading to the output buffer.
FETCh?
can be used to display the
measurement results in a number of different formats (for example, A/B
and B/A) without taking fresh data for each measurement.
CONFigure Examples
The following program segments show how to use the
READ?
command
and the
INITiate
and
FETCh?
commands with
CONFigure
to make
measurements.
It is advisable to read through these examples in order as they become
increasingly more detailed.
These examples give an overview of the
CONFigure
command. For further
information on the
CONFigure
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...