prog.book : chapter1.fm 85 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Status Reporting
HP E4418B/E4419B Programming Guide
1-85
Operation Status
The Operation Status group monitors conditions in the power meter’s
measurement process.
The Operation status model is shown in the pullout at the end of this
chapter.
The following bits in these registers are used by the power meter:
The condition bits are set and cleared under the following conditions:
Bit
Number
Decimal
Weight
Definition
0
1
CALibrating Summary
1 - 3
-
Not used
4
16
MEASuring Summary
5
32
Waiting for TRIGger Summary
6 - 9
-
Not used
10
1024
SENSe Summary
11
2048
Lower Limit Fail Summary
12
4096
Upper Limit Fail Summary
13 to 14
-
Not used
15
-
Not used (always 0)
Bit
Number
Meaning
EVENts Causing Bit Changes
0
CALibrating
This is a summary bit for the Operation
CALibrating Register.
SET:
At beginning of zeroing
(CALibration:ZERO:AUTO ONCE) and at the
beginning of calibration (
CALibration:AUTO
ONCE
). Also for the compound command/query
CALibration[:ALL]?
, this bit is set at the
beginning of the zero.
CLEARED:
At the end of zeroing or calibration.
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...