Power Meter Remote Operation
Zeroing and Calibrating the Power Meter
HP E4418B/E4419B Programming Guide
1-11
Zeroing and Calibrating the Power Meter
This section describes how to zero and calibrate the power meter.
The calibration and zeroing commands are overlapped commands refer to
“Using the Operation Complete Commands”, on page 1-89 to determine
when the commands are complete.
Zeroing
Zeroing adjusts the power meter’s specified channel for a zero power
reading with no power applied to the power sensor.
The command used to zero the power meter is:
CALibration[1|2]:ZERO:AUTO ONCE
The command assumes that there is no power being applied to the sensor.
It turns the power reference oscillator off, then after zeroing, returns the
power reference oscillator to the same state it was in prior to the command
being received. Zeroing takes approximately 10 seconds depending on the
type of power sensor being used.
When to Zero?
Zeroing of the power meter is recommended:
•
when a 5
0
C change in temperature occurs.
•
when you change the power sensor.
•
every 24 hours.
•
prior to measuring low level signals. For example, 10 dB above the
lowest specified power for your power sensor.
Calibration
Calibration sets the gain of the power meter using a 50 MHz 1 mW
calibrator as a traceable power reference. The power meter’s POWER REF
output or a suitable external reference is used as the signal source for
calibration. An essential part of calibrating is setting the correct reference
calibration factor for the power sensor you are using. The HP 8480 series
power sensors require you to set the reference calibration factor. The
HP E-series power sensors set the reference calibration factor
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...