prog.book : chapter1.fm 14 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-14
HP E4418B/E4419B Programming Guide
Making Measurements
The
MEASure?
and
CONFigure
commands provide the most
straight-forward method to program the power meter for measurements.
You can select the measurement’s expected power level, resolution and
with the HP E4419B the measurement type (that is single channel,
difference or ratio measurements) all in one command. The power meter
automatically presets other measurement parameters to default values as
shown in Table 1-3.
Table 1-3: MEASure? and CONFigure Preset States
An alternative method to program the power meter is to use the lower
level commands. The advantage of using the lower level commands over
the
CONFigure
command is that they give you more precise control of the
power meter. As shown in Table 1-3 the
CONFigure
command presets
various states in the power meter. It may be likely that you do not want to
preset these states. Refer to “Using the Lower Level Commands”, on
page 1-29 for further information.
Command
MEASure? and CONFigure
Setting
Trigger source
(
TRIGger:SOURce
)
Immediate
Filter
(
SENSe:AVERage:COUNt:AUTO
)
On
Filter state
(
SENSe:AVERage:STATe)
On
Trigger cycle
(
INITiate:CONTinuous
)
Off
Trigger Delay
(
TRIGger:DELay:AUTO
)
On
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...