prog.book : ch2meas.fm 8 Mon Feb 15 09:24:37 1999
MEASurement Instructions
CONFigure[1|2] Commands
2-8
HP E4418B/E4419B Programming Guide
CONFigure[1|2] Commands
The
CONFigure
commands are used on the specified window to set,
•
the expected power level value of the measurement.
•
the resolution of the window.
•
the channel(s) on which the measurement is to be made.
The
CONFigure
commands do not make the power measurement after
setting the configuration. Use
READ?
, or alternatively use
INITiate
followed by a
FETCh?
to make the measurement.
The
CONFigure
command also does the following to the channel(s) which
are in the specified window (the channel(s) in the window are specified in
the
<source list>
parameter):
INITiate:CONTinuous OFF
Sets the power meter to make one trigger
cycle when
INITiate
is sent.
TRIGger:SOURce IMMediate
Sets the power meter to make the
measurement immediately a trigger is
received.
TRIGger:DELay:AUTO ON
Enables automatic delay before making the
measurement.
SENSE:AVERage:COUNt:AUTO ON
Enables automatic filter length selection.
SENSE:AVERage:STATe ON
Enables averaging.
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...