CALibration Subsystem
CALibration[1|2]:AUTO <Boolean>|ONCE
HP E4418B/E4419B Programming Guide
4-7
CALibration[1|2]:AUTO <Boolean>|ONCE
This command is used to calibrate the specified channel when
ONCE
is
selected. The command assumes that the power sensor is connected to a
1 mW reference signal. The
0|OFF
parameter is only required for the
query response and is ignored in the command. If
1|ON
is selected, it
causes the error -224, “Illegal parameter value” to occur.
For the HP 8480 series power sensors the reference calibration factor used
during this calibration can be derived from either an active sensor
calibration table or the value entered using
CALibration:RCFactor
.
The actual value used is the one which was most recently set. That is, a
value entered using
CALibration:RCFactor
is overridden if a sensor
calibration table is subsequently selected and enabled. Conversely,
CALibration:RCFactor
overrides any reference calibration factor
previously set from a sensor calibration table. To determine the currently
set reference calibration factor use
CALibration:RCFactor?
The HP E-series power sensors have their sensor calibration tables stored
in EEPROM which means that the reference calibration factor is
automatically downloaded by the power meter.
Note
The power meter should be zeroed before calibration using the
CALibration:ZERO:AUTO ONCE
command.
Syntax
Example
Reset Condition
On reset, automatic calibration is disabled.
CAL1:AUTO ONCE
This command causes the power
meter to perform a calibration on
channel A.
CAL
1
2
:AUTO
0|OFF
ONCE
Space
?
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...