prog.book : chapter1.fm 43 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Using Frequency Dependent Offset Tables
HP E4418B/E4419B Programming Guide
1-43
Note
The legal frequency suffix multipliers are any of the IEEE suffix
multipliers, for example, KHZ, MHZ and GHZ. If no units are
specified the power meter assumes the data is Hz.
PCT is the only legal unit for offset factors and can be omitted.
The frequency and offset data must be within range. Refer to the
individual commands in Chapter 4 for their specified ranges.
Any offset values entered into the table should exclude the effect of
the sensor. Characterisation of the test setup independently of the
sensor allows the same table to be used with any sensor.
Ensure that the frequency points you use cover the frequency
range of the signals you want to measure. If you measure a signal
with a frequency outside the frequency range defined in the
frequency dependent offset table, then the power meter uses the
highest or lowest frequency point in the table to calculate the
offset.
To make subsequent editing of a frequency dependent offset table
simpler, it is recommended that you retain a copy of your data in a
program.
Listing the Frequency Dependent Offset Table Names
To list the frequency dependent offset tables currently stored in the power
meter, use the following command:
MEMory:CATalog:TABLe?
Note that
all
tables are listed; including sensor calibration tables.
The power meter returns the data in the form of two numeric parameters
and a string list representing all stored tables.
•
<numeric_value>,<numeric_value>{,<string>}
The first numeric parameter indicates the amount of memory, in
bytes, used for storage of tables. The second parameter indicates
the memory, in bytes, available for tables.
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...