prog.book : chapter1.fm 30 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Using Sensor Calibration Tables
1-30
HP E4418B/E4419B Programming Guide
Using Sensor Calibration Tables
This section applies to all HP 8480 series power sensors. It does not apply
to the HP E-series power sensors. The HP E-series power sensors have
their sensor calibration tables stored in EEPROM which allows frequency
and calibration factor data to be downloaded by the power meter
automatically.
This section describes how to use sensor calibration tables. Sensor
calibration tables are used to store the measurement calibration factors,
supplied with each power sensor, in the power meter. These calibration
factors are used to correct measurement results.
Overview
For the HP 8480 series power sensors there are two methods of providing
correction data to the power meter depending on the setting of the
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
command. If
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
OFF
the sensor
calibration tables are not used. To make a calibrated power measurement
when
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
OFF
, perform
the following steps:
1.
Zero and calibrate the power meter. Before carrying out the
calibration set the reference calibration factor for the power meter
you are using.
2.
Set the calibration factor to the value for the frequency of the
signal you want to measure.
3.
Make the measurement.
When
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe
is
ON
, the
sensor calibration tables are used, providing you with a quick and
convenient method for making power measurements at a range of
frequencies using one or more power sensors. Note that with the sensor
calibration table selected, the RCF from the table overides any value
previously set. The power meter is capable of storing 20 sensor calibration
tables of 80 frequency points each.
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...