prog.book : chapter1.fm 13 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Zeroing and Calibrating the Power Meter
HP E4418B/E4419B Programming Guide
1-13
Setting the Reference Calibration Factor
All the HP 8480 series power sensors require you to set the reference
calibration factor. The reference calibration factor can be set by:
•
entering the value into the power meter using the
CALibrate[1|2]:RCFactor
command.
•
selecting and enabling the sensor calibration table. The reference
calibration factor is automatically set by the power meter using
the reference calibration factor stored in the sensor calibration
table. See “Using Sensor Calibration Tables”, on page 1-30 for
further information.
Examples
a) To enter a reference calibration factor of 98.7% for channel A, you
should use the following command :
CAL:RCF 98.7PCT
This overides any
RCF
previously set by selecting a sensor
calibration table.
b) To automatically set the reference calibration factor, you have to
use a sensor calibration table as described in “Using Sensor
Calibration Tables”, on page 1-30. To select and enable the table
use the following commands:
[SENSe[1]]|SENSe2:CORRection:CSET1:SELect <string>
[SENSe[1]]|SENSe2:CORRection:CSET1:STATe ON
When the sensor calibration table is selected the
RCF
from the
table overides any value previously set.
Querying the Reference Calibration Factor
To determine the current reference calibration factor, use the following
command:
CALibration[1|2]:RCFactor?
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...