prog.book : chapter1.fm 16 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Making Measurements
1-16
HP E4418B/E4419B Programming Guide
Example 2 - Specifying the Source List Parameter
The
MEASure
command has three optional parameters, an expected power
value, a resolution and a source list. These parameters must be entered in
the specified order. If parameters are omitted, they will default from the
right. The parameter
DEFault
is used as a place holder.
The following example uses the source list parameter to specify the
measurement channel as channel A. The expected power and resolution
parameters are defaulted, leaving them at their current settings. The
measurement is carried out on the upper window.
The operation of the
MEAS1?
command when the source list parameter is
defaulted is described in the note on page 1-18.
Note
For the HP E4418B it is not necessary to specify a channel as only
one channel is available.
Example 3 - Specifying the Expected Power Parameter
The previous example details the three optional parameters which can be
used with the
MEASure?
command. The first optional parameter is used to
enter an expected power value. Entering this parameter is only relevant if
you are using an HP E-series power sensor. The value entered determines
which of the power sensor’s two ranges is used for the measurement. If the
current setting of the power sensor’s range is no longer valid for the new
measurement, specifying the expected power value decreases the time
taken to obtain a result.
The following example uses the expected value parameter to specify a
value of -50 dBm. This selects the power sensor’s lower range (refer to
“Range”, on page 1-47 for details of the range breaks). The resolution
parameter is defaulted, leaving it at its current setting. The source list
parameter specifies a channel B measurement. The measurement is
displayed on the lower window.
MEAS1? DEF,DEF,(@1)
specifies window
specifies channel
MEAS2? -50,DEF,(@2)
specifies window
specifies channel
specifies expected power value
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...