prog.book : ch4cal.fm 2 Mon Feb 15 09:24:37 1999
CALibration Subsystem
CALibration Subsystem
4-2
HP E4418B/E4419B Programming Guide
CALibration Subsystem
The
CALibration
command subsystem is used to zero and calibrate the
power meter. It is also used to set the reference calibration factor for the
power sensor which is being used.
The numeric suffix of the
CALibration
program mnemonic in the
CALibration
commands refers to a channel, that is
CALibration1
and
CALibration2
represent channel A and channel B respectively. If you are
using the single channel HP E4418B power meter the
CALibration2
commands are irrelevant and cause the error “Header suffix out of range.”
Zeroing and calibration of the power meter is recommended:
•
when a 5˚C change in temperature occurs.
•
when you change the power sensor.
•
every 24 hours.
•
prior to measuring low level signals. For example, 10 dB above the
lowest specified power for your sensor.
The
CALibration
commands, with the exception of
RCFactor
, are all
overlapped commands, that is, the power meter can continue parsing and
executing subsequent commands while initiated. Refer to “Using the
Operation Complete Commands”, on page 1-89 to determine when the
commands are complete.
Keyword
Parameter Form
Notes
Page
CALibration[1|2]
[:ALL]
[event; no query]
[:ALL]?
[event;query]
:AUTO
<Boolean>|ONCE
:ECONtrol
:STATe
<Boolean>
:RCALibration
<Boolean>
:RCFactor
<numeric_value>
[non-SCPI]
:ZERO
:AUTO
<Boolean>|ONCE
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...