prog.book : chapter1.fm 89 Mon Feb 15 09:24:37 1999
Power Meter Remote Operation
Status Reporting
HP E4418B/E4419B Programming Guide
1-89
Using the Operation Complete Commands
The
*OPC?
and
*OPC
commands allow you to maintain synchronization
between the computer and the power meter. The
*OPC?
query command
places an ASCII character 1 into the power meter’s output queue when all
pending power meter commands are complete. If your program reads this
response before continuing program execution, you can ensure
synchronization between one or more instruments and the computer.
The
*OPC
command sets bit 0 (Operation Complete) in the Standard
Event Status Register when all pending power meter operations are
complete. By enabling this bit to be reflected in the Status Register, you
can ensure synchronization using the HP-IB serial poll.
Procedure
•
Send a device clear message to clear the power meter’s output
buffer.
•
Clear the event registers with the
*CLS
(clear status) command.
•
Enable operation complete using the
*ESE 1
command (standard
event register).
•
Send the
*OPC?
(operation complete query) command and enter
the result to assure synchronization.
•
Send your programming command string, and place the
*OPC
(operation complete) command as the last command.
•
Use a serial poll to check to see when bit 5 (standard event) is set
in the status byte summary register. You could also configure the
power meter for an
SRQ
interrupt by sending
*SRE 32
(status byte
enable register, bit 5).
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...