Using
the
Measure
at
Marker
Function
The
measure-at-marker
function
uses
a
built-in
algorithm
to
simplify
the
process
of
taking
data.
Measurements
are
made
by
placing
the
marker
on
the
signal
of
interest
and
then
selecting
the
measure-at-marker
function.
The
algorithm:
1.
locates
the
signal
of
interest
2.
turns
the
scan
o
at
that
signal
3.
optimizes
the
measurement
dynamic
range
4.
measures
all
detectors
currently
selected
5.
displays
the
results
in
the
onscreen
marker
box
The
marker
box
remains
on
the
screen
until
the
marker
position
or
receiver
tuning
is
changed.
The
measure-at-marker
results
can
be
saved
to
the
internal
list
by
pressing
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADD
TO
LIST
.
The
contents
of
the
internal
list
can
be
viewed
and
remeasured,
as
desired.
Refer
to
Chapter
5
for
more
information.
The
dwell
time
used
for
each
detector
can
be
adjusted
using
the
dwell
time
function.
Refer
to
the
\Customizing
Standard
Settings"
section
earlier
in
this
chapter
for
more
information.
The
following
example
shows
how
to:
use
the
measure-at-marker
function
use
the
marker
box
store
signals
in
the
list
view
signals
in
the
list
1.
Set
the
receiver
to
a
known
state
by
pressing:
4
PRESET
5
2.
Turn
on
the
internal
calibrator
signal
to
use
as
a
test
signal
by
pressing:
4
INPUT
5
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
VIEW
CAL
ON
OFF
ON
3.
A
ctivate
the
marker
by
pressing:
4
TEST
5
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
TUNE
SPN
TUNE
4.
Rotate
the
knob
until
the
marker
is
on
the
peak
of
the
300
MHz
calibrator
signal.
5.
Measure
the
signal
at
the
marker
by
pressing:
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MEASURE
AT
MKR
The
measured
detector
values
are
displayed
on
the
screen
in
the
marker
box.
Refer
to
Figure
3-3.
6.
Save
the
measurement
to
the
list
by
pressing:
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
ADD
TO
LIST
3-16
Making
Compliance
Measurements
Содержание 85422E
Страница 14: ...7 13 Using 3rd Ord Meas Function 7 19 12 1 External Keyboard Template 12 8 Contents 7 ...
Страница 61: ......
Страница 81: ......
Страница 116: ...Figure 6 25 Using Linearity Check to Identify Distortion Products Making EMI Diagnostic Measurements 6 35 ...
Страница 137: ......
Страница 161: ......
Страница 183: ......
Страница 195: ......
Страница 210: ...If nonrecoverable system errors occur regularly contact your HP representative Error Messages 13 15 ...
Страница 211: ......
Страница 218: ...Customer Support 14 7 ...
Страница 219: ......
Страница 229: ......