Operation
54
Monitor on detection area
This screen enables to display the detection level distribution against the detection area at a
real time basis.
Fig. 54 Monitor on detection area
Laser operating days reset
This screen enables to clear the laser operating days.
After replacing the laser, clear the operating day in this screen.
Fig. 55 Laser operating days reset screen
Operation
1.
To clear the laser operating days, touch “YES”. When the clearing operation
completes, the screen returns to “Test mode menu”.
To close this screen without clearing, touch “NO”.
Note
When attempting a reset of the laser operating days frequently, it may disable to timely display the
message to prompt laser replacement.
Name
Description
Distribution of the
detection level
Divides the detection level in the section of 1 mA, displaying the rate of detection level in
each section to the entire detection area.
Displays 3 sections from the higher rate.
Distribution of the
detection level