Overview
8
Changing the setting
To change the setting, remove the front lid of display box. (Fig. 3), and operate the touch panel
on the LCD screen. Confirm the setting items in Table 4.
Fig. 3 Display box front view
Table 4 Setting items
Note
If the front cover of the display box is removed for the access to the touch panel to change settings or
reset alarms, be sure to place the cover and fasten four screws with a torque wrench (0.6 N•m) for final
fitting. Improper torque or offset O-ring from the groove will lead water inside the detector and the
instruments will be damaged.
Item
Menu button
Description
Detection level setting
OIL LEVEL
Sets the threshold of the detection level
Detection area setting
AREA LEVEL
Sets the threshold of the detection area
Continuous judgment time
setting
CONT.TIME
Sets the required continuous judgment time for the
alarm output
Contact output setting
RELAY OUT
Selection of oil film warning contact output (alarm
or warning)
Setting of contact output at hold time
Lower limit error judgment
setting
LEVEL ERR.
Sets ON/OFF of lower limit error
Multi-Trans function setting
MULTI-TRANS
Sets ON/OFF of Multi-Trans function
Upper/lower limit setting
OVER LEVEL
Sets threshold at upper/lower limit of detection level
Time/date setting
DATE
Sets date and time
Contrast adjustment
CONTRAST
Adjusts contrast of screen
Color setting
COLOR
Changes the color of screen and characters
Laser replacement
Contact output time setting
LM ALARM TIME
Setting the time to output the system error contact,
after the laser replacement time (approx. 1460 days)
has passed.
Remove bolts of 4
M3 A from the front side of display box