Overview
5
Display examples of detection result
Oil film alarm status
The state in which the oil film alarm is triggered due to trace volume of oil when the area
detection function is invalid.
Monitoring of detection level
The detection area is checked by the detection level monitor in the Test mode.
Setting of detection area
An oil film of 20
is continuously suspended on the detection area; the threshold of alarm is
set to 50
, and the threshold of warning is set to 30
.
Monitoring status after setting change
Check that the oil film alarm indication disappears and the status mode turns into NORMAL
RUN.
Note
Use the area detection function with the distance from the detection area to the detector within 2
m.
The detection area depends on the shape and the position of oil film.
If the area detection function is activated in a location with ripples, it might not detect oil film.