![horiba LO-300 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/horiba/lo-300/lo-300_instruction-manual_2162291024.webp)
Overview
10
Other functions
Water level automatic adjustment function
The water level automatic adjustment function is provided for maintenance. When the
receiving light adjustment at the water level is needed because of the optical part
replacement, automatic adjustment is available by touch panel operation on the monitor of
display box.
Reference
For water level automatic adjustment, refer to “ Water level adjustment ” (page 53).
History display function
LO-300 can save up to 100 sets of history information.
The history information is as follows:
Oil film detection (alarm)
Motor stop, frequency of 40 Hz or less, or luminescence stop
Oil film detection (warning)
Motor rotation error
Upper limit error
Notification of laser replacement
Lower limit error
Normal return (from oil film alarm, oil film warning, upper limit error, lower limit error)
Temperature error
Reference
To confirm the history information, refer to “ History ” (page 49).
Laser replacement display function
In the LO-300, when the working days reaches the laser replacement time (approx. 1450
days), a message prompting laser exchange is displayed on the monitoring monitor screen of
the display box. Also, when the laser replacement time (approx. 1460 days) has elapsed, a
system error contact will be output. Oil film monitoring continues, but if this message is
displayed, please replace the laser earlier. For system error contact release, please refer to
“ Releasing contact output ” (page 21).
Reference
For the laser replacement display function, refer to “ Laser replacement message ” (page 19) or “
Laser operating days reset ” (page 54).